The plate for output is a critical accessory whose condition significantly affects the performance of the RE therapeutic device, and itis also a consumable item. Please replace the Plate for output at appropriate intervals and handle it properly.
Replacement of plate for output
Handling of plate for output
- A damp condition of the plate for output will accelerate its deterioration.
-
If sweat gets on the plate for output during use, dry it afterward.
If the cover of the plate for output becomes wet, replace it with a dry cover. - There is no distinction between the front and back of the plate for output and the plate for ground.
-
By alternating between both sides, moisture and deterioration can be reduced.
- Do not use the plate for output and the plate for ground in a curved manner.
-
Load is applied to the curved part, causing further deterioration.
Disconnection of cable of plate for output
If force is continuously applied in the same direction to the plate for output and the plate for ground cables, especially to the connection between the plate and the cable, or if a large force is applied even for a short period, the cable may break as shown in the figure below.
To prevent cable breakage, please pay attention to the following:
- Please don’t bend the cable sharply, especially around the connection between the cable and the plate, or use the cable in a bent state.
- Please don’t place heavy objects on the cable.
- Please don’t twist the cable or pull on the connection between the cable and the plate.
- When storing cables, only roll them up moderately small (do not bundle them with a small diameter).
If the surface of the cable or plate is torn or damaged even slightly, do not use it; replace it with a new one.